1.Equipped with 80 bit auto-input sampler, put the sample into the sampler, start button to start measuring until the end of measuring, completely automated by software control. 1.Equipped with 80 bit auto-input sampler, put the sample into the sampler, start button to start measuring until the end of measuring, completely automated by software control.
2.The host built-in air pump, provides stable, priceless cleaning air for carrier gas. 2.The host built-in air pump, provides stable, priceless cleaning air for carrier gas. 2.The host built-in air pump, provides stable, priceless cleaning air for carrier gas.
3.Using permutation pump self-suction rapid quantitative sampling and reagents; 3.Using permutation pump self-suction rapid quantitative sampling and reagents; 3.Using permutation pump self-suction rapid quantitative sampling and reagents;
4. Nitrate nitrogen ani total nitrogen ye tɛmɛ tɛmɛ tɛmɛ tɛmɛ tɛmɛ tɛmɛ tɛmɛ tɛmɛ tɛmɛ tɛmɛ tɛmɛ tɛmɛ tɛmɛ tɛmɛ tɛmɛ
5.Single sample measurement, from sampling → measuring results → measuring the next sample with ≤50s.
Teknik Indikator
1.Stability: 30min zero baseline drift ≤ ± 0.0001Abs, continuous stable working time greater than 12 hours.
2. Signal noise ratio: zero baseline noise ≤± 0.0002Abs;
3.Detection limit: Lower than HJ/T195~200 (2005) standard method, test detection limit is below:
Nitrogen ammonia 0.005mg/L, Nitrogen nitrite 0.003mg/L, Nitrogen nitrite 0.001mg/L, Nitrogen total 0.010mg/L,
Nitrogen 0.01 mg/L, sulfide 0.002 mg/L, mercury 0.001 maximum/L.
4. Kalibration curve relevance: r>0.999
5. Luɔi: ≤±3%
6. Kuen: RSD ((n=6) ≤3%
7. Kuɛl: 0.01 ~ 100mg/L